SLE 2000 artificial lung ventilation apparatus for newborns
Function
The SLE 2000 artificial lung ventilation apparatus is designed for controlled ventilation of the lungs in newborns and children in the process of resuscitation in the conditions of clinical hospitals, maternity hospitals, research institutes of the children's profile.
The unique technology used in the device eliminates the need for expiration valves, diaphragms and provides ventilation for patients weighing from 500 g to 20 kg.
The SLE 2000 artificial lung ventilation apparatus is a fan with pressure control and time switching (time-cyclic). The SLE 2000 artificial lung ventilation apparatus has all the necessary parameters for artificial ventilation in mature and premature neonates.
The apparatus provides support for the constant pressure wave form at all respiratory frequencies, has an alarm system, a built-in oxygen analyzer with a digital indicator and the option of exhaled gases filtering.
Advantages
Ventilation Modes
Specifications
Ventilation Modes |
|
BRP (breathing rhythm) | 1 - 125 (126 - 250 with the Z0004 option) |
Maximum inhalation time | 0.1 - 3.0 sec. (0.01 - 0.3 seconds with the Z0004 option) |
Inhalation/expiration rate (I:E) | from 9.9: 1 to 1:99 |
Manual inhalation | single breath in CPAP, CMV, PTV, SIMV modes |
PTV Sensitivity | allowable maximum |
Oxygen concentration | 21 - 100% O2 ± 3% |
Pressure in CPAP mode | 0 - 15 mbar |
Inhalation pressure | 0 - 60 mbar |
Alarm disabling for 60 seconds | |
Reset alarms | resets all alarms (excluding system errors) |
Light indicators | |
Network | green light, the network is on |
System error | red signal, an error in the main processor |
Backup Trigger | green signal - the machine provides breathing |
Digital | |
Breathing frequency monitor | inhalations per minute |
Inhalation time monitor | inhalation time |
Breathing frequency monitor | inhalation /expiration value |
Oxygen concentration | from 21 to 100% |
Pressure | from 0 to 62 mbar |
Calibration | |
Pressure is calibrated | from 0 to 60 mbar plus PEEP level |
Alarms | |
CPAP | visual and intermittent sound signals |
PIP / cycle error / low readings | visual and intermittent sound signals |
Blockage of air or leakage | visual and intermittent sound signals |
Loss of main power | battery powered, buzzer |
Loss of air or oxygen supply | pneumatic, sounds from the blender |
System error | visual and intermittent sound signals |
Complete set, dimensions and weight | |
Input air and oxygen pressure | 3 - 5 bar |
Voltage | 100-120 V, 50/60 Hz; 220-240 V, 50/60 Hz |
Power | 20 VA |
Circuit breakers | 100-120 V = T500 mA 220-240 V = T200 mA |
Protection | Class I Type B |
Dimensions, only the apparatus | 370mm (L) x 310mm (H) x 320mm (W) |
Rack height | 12V |
Weight | 10 kg |