AED-A15 Automatic External Defibrillator


Function

AED is intended for defibrillation of patients with sudden cardiac arrest due to the development of ventricular fibrillation and ventricular tachycardia in medical and pre-medical care, as well as for public and home use.


Advantages

  • Compact and lightweight (weight 2.65 kg)
  • Simple and convenient: defibrillation in 4 actions
  • Visual and sound accompaniment of operator actions
  • Self-test mode (every 24 hours, once a week, once a month)
  • Disposable electrodes are placed in the storage section (under the defibrillator cover)
  • The patient mode switch allows the use of a single set of electrodes both for adults and children
  • Service life - 5 years


Delivery set


Name
Amount, pcs.
1.
Automatic external defibrillator А15, including:
1
1.1
Disposable electrodes
1
1.2
Power supply battery
1
1.3
Antiseptic alcohol sterile wipe
1
1.4
Shaving stick
1
1.5
Soft carrying bag
1
1.6
Memory card SD
1
2.
Documents
1


Specifications

Discharge
Pulse shape
Bipolar (bi-phase) trapezoidal with compensation of the patient's impedance
Discharge energy
Adult mode: (170 – 200) J (±15 %)
Child mode: (45 – 50) J (±15 %)
Mode
Semi-automatic
Electrocardiogram
Leads
II (RA, LL)
Range of chest resistance values
(25 – 200) Ohm
Heart rate
(20 - 300) bpm
Heart Rate Error
1 bpm
Recognition
Ventricular fibrillation (VF)
≥0.2 mV
Ventricular tachycardia (VT)
≥160 bpm
Connection of electrodes
Sound confirmation
Frequency
(0.5 – 30) Hz
Control
Basic
Lid opening button, Patient mode switch, "Discharge" button
Indicators
Visual
Action indicator panel, status display (status of AED A15, batteries, temperatures), LED indicators (patient mode switch)
Sound
Speaker (voice prompts):
Sound intensity at a distance of 1 m from the AED A15, not more than

Buzzer (beeps (CPR)):
Sound intensity at a distance of 1 m from the AED A15, not more than
Acoustic zoomer frequency 2,300 ± 300 Hz

Physical characteristics

Overall dimensions
Not more than (240 × 294 × 95) mm (W × H × D)
Weight
No more than 2.65 kg with battery without electrodes

Application environment

Operation / Standby mode
Temperature of
(0…43) °C
Relative humidity
(5 – 95) % (without moisture condensation)
Working height
(0 – 4,575) m
Allowable drop height
1 m
Ingress protection
IP54 (IEC60529)
Note: The standby mode informs about the periodic start of the self-test with the battery installed in the AND A15
Storage (in transport packaging)
Temperature of
(20…60) °C
Relative humidity
(5 – 95) % (without moisture condensation)
Working height
(0 – 12,192) m
Self-test
Temperature of
(0…43) °C
Periodicity
Every 24 hours, once per week, once per month
Results
The status display displays "O", or "X"
Note: The self-test can only be started when the battery is installed

Backup and data transfer

Standard
SD memory card, IR port

Disposable electrodes

Adult / Pediatric electrodes
Shelf life
2 years from the date of manufacture
Electrodes
Disposable electrodes
Location:
Adults: Anterolateral
Pediatric: Anteroposterior
Minimum number of discharges at maximum energy
10 energy discharges 200 J
Total electrode area
120 см2 (±5 %)
Minimum active (gel-coated) electrode area
80 см2
Width of adhesive electrode ring
11 mm (±5 %)
Electrodes cable length
Not less than 2 m
Environmental conditions
Temperature of
Operation: (0…43) °C
Storage: (0…43) °C
Relative humidity
(5 – 95) % (without moisture condensation)

Battery Power

Type
LiMnO2, non-rechargeable, with a long life
Voltage / Capacity
15 V, 4,200 mAh
Shelf life (in original packaging)
2 years from the date of manufacture
Standby time (when connected to AND A15)
5 years from the date of manufacture
Performance
At least 200 200 J energy discharges (excluding the period between CPR and defibrillation) or 10 hours of operation in monitoring mode at an ambient temperature of 20 °C
Environmental conditions
Temperature of
Operation: (0…43) °C
Storage: (0…43) °C
Relative humidity
(5 – 95) % (without moisture condensation)